Directions for Use
Chapter 5 Performing ECG Tests
59
Figure 39. “Auto ECG Post-Print” Screen
8.
Press the desired softkey.
9.
Press Exit to complete patient testing.
Softkey
Effect
Print Copy
Prints a copy of the test.
To learn how to print multiple copies of all tests automatically, see
“Reviewing the
Interpretation and Copy Settings for Auto Reports”
on page 43.
Edit Test
Brings up the Edit Test – Patient Data screen. From here you can edit or confirm the
patient data and interpretation statements. See
“To Edit or Confirm a Test Directly
After Printing”
on page 61.
Manual Save
Saves the test to the electrocardiograph’s test directory. See
“Managing Saved Tests
In the Test Directory”
on page 70.
This Manual Save softkey appears only when Auto Save is disabled. For more on
Auto Save, see
“Reviewing the Miscellaneous ECG Settings”
on page 48.
Manual Send
(May appear in place of Manual Save in
Figure 4
on page 14.) Displays two softkey
options:
•
Memory Card
An SD memory card or removable USB storage device must be in place.
•
Workstation
Either wireless communications must be configured or a USB cable must connect
the CardioPerfect workstation to the electrocardiograph’s Com port B (
).
This Manual Send softkey appears only when Auto Save is enabled. For more on
Auto Save, see
“Reviewing the Miscellaneous ECG Settings”
on page 48.
Repeat Test
Lets you repeat a test for the current patient.
Exit
The ECG Preview screen appears if all leads are connected to the patient.
WARNING Failure to follow these instructions could incorrectly associate an
ECG test with the wrong patient.
Copy
Doe, Jane
Edit
Test
Manual
Save
Repeat
Test
Exit
9:17AM Oct 16 05
Auto ECG Post-Print
I
V
aVR
Softkeys for choosing
post-printing options
HR 120
Note
Print Copy is not available for Stat Auto ECG, or if
there is no Patient Name or ID entered for Auto