Troubleshooting – Welch Allyn PROPLUS 2650KGWA Health-o-Meter Professional Scale - User Manual User Manual

Page 11

Advertising
background image

10

TROUBLESHOOTING

Refer to the following instructions to check and correct any failure before contacting Customer Service.

SYMPTOM

POSSIBLE CAUSE

CORRECTIVE ACTION

Scale does not turn on

1. Dead Battery
2. Faulty electrical outlet
3. Bad power supply

1. Replace batteries
2. Use a different outlet
3. Replace adapter

Questionable weight or
the scale does not zero

1. External object

interfering with the scale

1. Remove interfering object

from the scale

2. The display did not show

”0.0” before weighing

2. Ask the patient to get off the

scale, zero the scale and
begin weighing process
again

3. Scale is not placed on a

level floor

3. Place the scale on a level

floor and begin weighing
process again

4. Scale is out of calibration 4. Check weight with known

weight value

5. Improper Tare

5. Place the item to be tared

on the scale. Press
REWEIGH. Once the weight
of the item is displayed,
press TARE. Place the
patient and the tared item
back on the scale. Press
REWEIGH again

Weighing is performed but
the display shows “weigh”
and “reweigh” every few
seconds; the weighing
process takes too long and
no weight is displayed.

The patient is not standing
still

Ask the patient to stand still or
you can change to live weight
setting

The display shows
“Overload” message

The load on the scale
exceeds the capacity
(400 kg)

Remove the excess weight and
use the scale according to its
limits

The display shows
“LOW BAT” message

The batteries are empty

Replace batteries according to
instructions

The display shows
“Load Cell Error” message

There is a problem with one
or more load cells or the load
cell cable is disconnected.

Check load cell cable connection
at the display and platform
assembly ports. If the problem is
not corrected, refer to qualified
service personnel to replace the
defective load cell