Detailed views, Troubleshooting (continued), Error list error no. definition check/replace – Welch Allyn Health-o-Meter Professional Scales - User Manual User Manual
Page 10
9
TROUBLESHOOTING (CONTINUED)
ERROR LIST
ERROR
NO.
DEFINITION
CHECK/REPLACE
Err 1
Mathematical mistake; division by zero.
Cannot calculate calibration factor
Will show when trying to calibrate with no
calibration weight
Err 2
Weight loaded too low. Cannot calibrate
While in calibration procedure add more weight in
load mode
Err 3
Unstable weight. Cannot calibrate
Check the load cells mechanical surroundings and
see nothing touches them and the cables are well
connected
Err 5
Calibration disabled
Toggle the calibration switch to enable
Err 6
High saturation state (high A/D)
The load cell is not connected correctly. Check
the connection, cables, and mechanical
connection. If the problem persists, replace the
set of load cells
Err 7
Low saturation state (low A/D)
See Err 6