This Quick reference card provides steps for initiating testing with the Welch Allyn

®

OAE Screener. Refer to the Welch Allyn OAE Screener

Directions for use for complete instructions and information.

Buttons and menus

Navigation buttons

Down button

• Single press: turn on power
• Single press: enter the Protocol

menu

• Press and advance to the Date

menu. Press and hold to enter

device settings menus

Right button

• Main menu: start a test on the right ear
• All menus: advance through options in the selected menu

Left button

• Main menu: start a test on the left ear
• All menus: advance through options in the selected menu

Up button

• Single press: move up one level in the menu structure
• Single press: on the Main menu, move to the Print menu at the

completion of a test

Menu structure

1. Power up screen: displays the software and firmware version of the

screener

2. Main menu: displays the selected test protocol and

L and R for

testing each ear; allows advancing to and selecting stored protocols

3. Date/Time menu: displays the date and time and allows changes

using the left and right control panel buttons

4. System menu: displays device operation settings; allows changes

using the left and right control panel buttons

5. Protocol menu: displays the protocol settings and allows changes

using the left and right control panel buttons. Changes to settings

are available on the screener.

OAE hearing screener

Connect cable to screener

1. With the screener turned off, insert the HDMI connector firmly

into the socket on the top of the screener. The plug fits only in one

direction.

Attach the probe tube

1. Insert a new probe tube into the probe head until it is fully seated.

A properly inserted probe tube snaps securely into place when it is

fully seated in the probe head.

2. To remove a probe tube, grasp the tube and pull gently away from

the probe head with a slight twist.
It is not necessary to replace the probe tube with each eartip; the

tube is reusable as long as the probe tube is clear.

Attach an eartip

WARNING Safety risk. The eartips are disposable and

must be replaced after each patient.

1. Select the appropriate-sized eartip.
2. Push and slightly twist the eartip onto the probe tube.
3. Ensure that the eartip is fully seated on the

probe. There should be no gaps between the

eartip and the collar of the probe tube.

Select the Protocol

The OAE screener provides two (2) protocol options: DP 4s–testing at

a maximum time of 4 seconds per frequency and DP 2s–testing at a

maximum time of 2 seconds per frequency.
1. Press the

Down control panel button on the Main menu. The

Protocol menu appears.

2. Press the

Left or Right control panel button to select the desired

protocol.

3. Press the

Up control panel button to return to the Main menu and

begin testing.

39403

Perform a screening test

1. Insert the probe with eartip into the ear canal.
2. Power on the screener by pressing the

Down control panel button.

The Main menu appears.

3. Press the

Left or Right control panel button to select the test ear.

The screener performs an automatic probe and calibration test.

The screening automatically starts. When the test is complete, the

screener returns to the Main menu and displays either a

Pass, Refer,

Noisy, or No Seal or Fit Err message.
4. If either the

Noisy or No Seal or Fit Err message appears, readjust

the probe placement, and then retest.

5. To review test results, press the

Down control panel button before

testing the opposite ear. If you do not review the results, they will

be unavailable for viewing until the results are either printed or

transferred to the data manager.

6. To proceed to the next screening test, insert the probe into the

opposite ear. Press the

Left or Right control panel button to test

the opposite ear.

Print test results

You must have an optional Bluetooth-enabled thermal printer to print

from the screener. Refer to the Welch Allyn OAE Screener

Directions for use.
1. Confirm that the screener is paired with the printer.
2. Confirm that the printer is on and ready for printing.
3. From the Main menu, press the

Up control panel button. The

Connection screen appears.

4. Press the

Left or Right control panel button to connect to the

printer. All stored results are printed.

© 2014, Welch Allyn, Inc. Mat. No. 105763, 80019149 Ver. A