Lead alert or square wave

A dot might be flashing on the lead-status screen. Or one or more leads might appear as

a square wave.

Causes
•

Electrode contact might be poor.

•

A lead might be loose.

•

A lead might be defective.

Actions
•

Replace the electrode.

•

Verify that the patient’s skin has been properly prepared.

•

Verify that electrodes have been properly stored and handled.

•

Replace the patient cable.

System problems

The electrocardiograph won't turn on when it is plugged in

Causes
•

The AC power connection is faulty.

•

An AC fuse is blown.

•

There is no AC power.

Actions
•

Check the AC power source.

•

Check the AC fuses.

The electrocardiograph won't turn on when it is unplugged

Causes
•

The battery is disconnected or incorrectly connected.

•

The battery is low, not charging, depleted, or bad.

Actions
•

Check the battery connections.

•

Recharge the battery.

•

Replace the battery.

The electrocardiograph shuts down during printing

Causes
•

The battery is low or bad.

Actions