Recharge the battery.

Replace the battery.

The electrocardiograph prints fewer than 10 reports on a full battery

charge

Causes

The battery is degraded.

Actions

Replace the battery.

The electrocardiograph does not respond when you press buttons or

touch the screen

Causes

The electrocardiograph is frozen.

Actions

Reset the electrocardiograph by pressing and holding the power button for at least

six seconds until the screen goes blank. Press the power button again. The

electrocardiograph goes through some diagnostic tests that cause it to take longer

than usual to power up.

Note

For more troubleshooting guidelines, see the service manual.

CAUTION The service manual is for use only by qualified service
personnel who understand technical English.