Troubleshooting, Lead-quality problems, Artifact” message on the screen – Welch Allyn CP150 12-Lead Resting Electrocardiograph, Software Version 2.00.XX - Quick Reference Guide User Manual

Page 13: Wandering baseline

Advertising
background image

Troubleshooting

Lead-quality problems

“Artifact” message on the screen

Artifact is signal distortion that makes it difficult to accurately discern the waveform

morphology.

Causes
•

The patient was moving.

•

The patient was shivering.

•

There is electrical interference.

Actions

See actions for wandering baseline, muscle tremor, and AC interference.

Wandering baseline

Wandering baseline is an upward and downward fluctuation of the waveforms.

Causes
•

Electrodes are dirty, corroded, loose, or positioned on bony areas.

•

The electrode gel is insufficient or dried.

•

The patient has oily skin or used body lotions.

•

Rising and falling of chest during rapid or apprehensive breathing.

Actions
•

Clean the patient’s skin with alcohol or acetone.

•

Reposition or replace the electrodes.

•

Verify that the patient is comfortable, warm, and relaxed.

•

If wandering baseline persists, turn the baseline filter on.