Muscle tremor

Causes
•

The patient is uncomfortable, tense, nervous.

•

The patient is cold and shivering.

•

The exam bed is too narrow or short to comfortably support arms and legs.

•

The arm or leg electrode straps are too tight.

Actions
•

Verify that the patient is comfortable, warm, and relaxed.

•

Check all electrode contacts.

•

If interference persists, turn the muscle-tremor filter on. If interference still persists,

the problem is probably electrical in nature. See the suggestions for reducing AC

interference (in a related troubleshooting tip).

AC interference

AC interference superimposes even-peaked, regular voltage on the waveforms.

Causes
•

The patient or technician was touching an electrode during recording.

•

The patient was touching a metal part of an exam table or bed.

•

A lead wire, patient cable, or power cord are broken.

•

Electrical devices in the immediate area, or lighting, or wiring concealed in walls or

floors are interfering.

•

An electrical outlet is improperly grounded.

•

The AC filter is turned off or set incorrectly.

Actions
•

Verify that the patient is not touching any metal.

•

Verify that the AC power cable is not touching the patient cable.

•

Verify that the proper AC filter is selected.

•

If interference persists, unplug the electrocardiograph from AC power and run it on

the battery. If this solves the problem, you’ll know that the noise was introduced

through the power line.

•

If interference still persists, the noise may be caused by other equipment in the

room or by poorly grounded power lines. Try moving to another room.

12 Troubleshooting