Installation instructions, Spiroperfect – Welch Allyn SpiroPerfect Spirometry Quick Start - Quick Reference Guide User Manual
Page 3

•
Use the USB extension cable supplied
with the sensor to lessen chance of
USB connector damage
•
Connect the pressure tubing to the
sensor and the other end of the
tubing to a flow transducer.
•
Connect the USB sensor to a USB port on
your computer.
1 Install Software and USB Drivers
2 Connect USB Dongle to CPU
Each lot of flow transducers has a specific Calibration code.
Make sure to update these in the SW when you receive a
new lot of flow transducers.
The Lot number and Calibration Code are located on the
outside packaging label of the flow transducer box. Update
this information in the pre-calibration window or in the
Recording tab of the spiro settings.
3 Update Calibration Code
•
This quick start guide is intended for use with CardioPerfect
Workstation Version 1.6.0 or higher.
•
For complete directions for use and warnings, please
consult your user manual located on the SW CD.
SPIROPERFECT
Q U I C K S T A R T G U I D E