Troubleshooting Guide
PROBLEM
SOLUTION
No powder or air from the
handpiece nozzle.
??
The handpiece is clogged. Use an orthowire to clear the nozzle. Disconnect
the handpiece from the handpiece hose and blow air through it.
Air from the handpiece but no
powder flow.
??
No Powder
??
Blown Pinch Tube
??
Clogged orifice plate. Empty the powder from the canister. With an ortho
wire, gently move the wire back and forth in the small holes in the bottom of the
Powder Canister.
??
Pinch Valve Piston stuck
Air and powder from the
handpiece flow for only a few
seconds then stop.
??
A pinch tube is sticking. Replace the pinch tube.
??
Pinch Valve Piston stuck
Air and powder flow
continuously from handpiece
even when in the Standby
mode.
??
Change pinch tube.
??
Shoe block cover has become scarred. Replace the shoe block cover.
Powder inside the unit
chassis.
??
Blown Pinch Tube
??
Hole in Powder tubing
??
Shoe block cover has become scarred. Replace the shoe block cover.
Unit Plugged into Water
Contact Welch Allyn Service center for unit Repair