Setup – Welch Allyn Connex Vital Signs Monitor PMP Tab Instructions - Quick Reference Guide User Manual
Page 16

Part number
Description
Quantit
y
1/8 tube
1
Setup
1. Connect one end of a stereo jack plug to the PMP service test box and the opposite end to the
device.
2. Connect a USB cable to the PMP service test box; connect the opposite end to a PC.
3. Connect a mini B USB cable to the PMP service test box and the opposite end to the device
(comms port 1).
4. Connect the blood pressure hose to the NIBP module and the test volume.
•
Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite
end to the square fitting connector.
•
Connect another piece of tubing to the square fitting connector; connect the opposite end
to the calibrated pressure meter.
•
Connect another piece of tubing to the square fitting connector; connect the opposite end
to the test volume.
•
Connect the last piece of tubing to the square fitting connector; connect the opposite end
to the "Y" fitting on the two-tube blood pressure connector.
12 PMP calibration and verification setup