Setup – Welch Allyn Calibration/Verification Braun Thermoscan - Quick Reference Guide User Manual
Page 3
Setup
Connect the Braun ThermoScan module to perform the verification process.
Test equipment part numbers
Use the equipment detailed in the table below for the Braun ThermoScan verification. Contact Welch
Allyn customer service to purchase or locate specific parts.
Setup
C.
Retake measurement.
System error.
Wait one minute while the
thermometer turns off.
Turn thermometer on again.
If error persists, reset the
thermometer by removing, then
replacing the batteries.
Contact Welch Allyn customer
service.
Part number Description
Quantity
01802-110
Tester, Calibration, 9600 Plus (North America,
EMEA)
1
01802-119
Tester, Calibration, 9600 Plus (Latin America,
Asia Pacific)
1
01802-220 Tester,
Calibration, 9600 Plus (UK)
1
4500-925
USB 2.0/5-pin Mini-B cable, gold, 6 feet
1
715701
VSM6000 service test box
1