Welch Allyn VSM300 Addendum To - Quick Reference Guide User Manual

Addendum, How changing the patient type affects map defaults, Monitor behavior during equipment alerts

Advertising
background image

Welch Allyn Vital Signs Monitor

1

Addendum to the Directions for Use

810-1881-00 Rev A

Addendum

This addendum contains the following information:

How changing the patient type affects MAP defaults.

This information is not covered

elsewhere in the DFU.

Monitor behavior during equipment alerts.

This information supercedes the information

presented in the DFU

(

"Recoverable Alerts" on page 68

).

How Changing the Patient Type Affects MAP Defaults

When you cycle power to the monitor, the monitor stores all current settings before shutting
down, and then uses these settings when it powers up again. Note that this does not affect
the factory defaults.

Whenever you enable or disable MAP (mean arterioral pressure) for a given patient type
(Adult, Pediatric, or Neonatal), the current MAP enabled/disabled setting becomes the

default power-up setting

for that patient type.

For example: If the monitor is set to the Neonatal patient type and you set MAP Disabled,

MAP Disabled

becomes the default setting for neonatal patients.

Monitor Behavior During Equipment Alerts

Recoverable Temperature, NIBP, or SpO

2

Alert — Not Escalated

Most recoverable equipment alerts are not escalated to the level of patient alarms. When an
unescalated alert occurs, take the necessary steps to correct the equipment problem and then
resume monitoring the patient.

For an unescalated equipment alert for Temperature, NIBP, or SpO

2

, the monitor does the

following:

beeps once

displays the appropriate error code (Cxx) in the appropriate window (Temperature,
SYS, DIA, or SpO

2

)