Recoverable spo2 alert - escalated, Recoverable nibp alert - escalated, Battery alert – Welch Allyn VSM300 Addendum To - Quick Reference Guide User Manual
Page 2: Recoverable spo, Alert — escalated, Recoverable nibp alert — escalated
Advertising

810-1881-00 Rev A
Addendum to the Directions for Use
2
Welch Allyn Vital Signs Monitor
Recoverable SpO
2
Alert — Escalated
An SpO
2
equipment alert is immediately escalated to the level of a patient alarm if it occurs
when
both
of the following conditions are true:
•
SpO
2
monitoring has begun and the monitor has recorded an SpO
2
measurement
•
An SpO
2
or Pulse Rate alarm limit has been set
Recoverable NIBP Alert — Escalated
An NIBP equipment alert is escalated to the level of a patient alarm whenever two
consecutive NIBP alerts occur while the monitor is taking automatic NIBP/PR
measurements and alarms are enabled.
Battery Alert
A battery alert is described in the DFU (
"Battery Low Warning"
and
"Battery Failure" on
page 87
).
Advertising