Nextteq Chloroform User Manual

Page 2

Advertising
background image

**Shelf Life: Please refer to the Validity Date printed on the box of tube.
**Store the tube in the refrigerator to keep at 10

o

C (50

o

F) or below.


CORRECTION FOR TEMPERATURE, HUMIDITY AND PRESSURE:

Temperature:

Correct for temperature with the table below.

Temperature

o

C

(

o

F)

0

(32)

5

(41)

10

(50)

15

(59)

20

(68)

25

(77)

30

(86)

35

(95)

Correction Factor

2.3

1.8

1.4

1.1

1.0

1.0

1.0

1.0

Humidity:

No correction is required.

Pressure:

To correct for pressure, use the formula below.


Tube Reading (ppm

)

x 1013 (hPa)

Atmospheric Pressure (hPa)


MEASUREMENT PROCEDURE:
1. For checking the leakage of the pump, insert a freshly sealed detector tube into pump.

Follow the instructions provided with the pump operating manual.

2. Break the tips off a fresh primary tube and analyser tube using the tube tip breaker of the

pump.

3. Connect

©

marked ends with rubber tubing after breaking each end.

4. Insert the analyser tube securely into pump inlet with arrow (

) on the tube pointing

toward the pump.

5. Make certain the pump handle is all the way in. Align guide marks on pump body with the

guide marks on the handle.

6. Pull the handle all the way out until it locks at one pump stroke (100 mL). Wait two minutes

and confirm the completion of the sampling. Repeat three more times.

7. Read concentration level at the interface where the stained reagent meets the unstained

reagent.

8. If necessary, multiply the reading by the correction factors of the pump strokes and

atmospheric pressure respectively.


INTERFERENCES:

Substance

Concentration

Interference

Interference gas only

Chlorine

6ppm +

Pale purple

Bromine

4.5ppm

No No

Iodine

4.5ppm No

No

1,1,1-trichloroethane

+ Pale

purple

Trichlorethylene

+

Pale purple

The table of this interference gases primarily expresses the interference of each coexisting gas in
the gas concentration range, that is equivalent to the gas concentration. Therefore, the test result
may show positive result due to other substances not listed in the table. If more information is
needed, please contact us or our distributors in your territory.

DANGEROUS AND HAZARDOUS PROPERTIES:
Threshold Limit Value-Time Weighted Average by ACGIH (2009): 10 ppm
INSTRUCTIONS ON DISPOSAL:

Reagent of the tube does not use toxic substance. When disposing the tube regardless of
whether it has been used or not, follow the rules and regulations of the local government.

WARRANTY:
If you have any questions regarding gas detection and quality of the tubes, please
feel free to contact your Gastec representatives.

Manufacturer: Gastec Corporation

IM00137LLE1

8-8-6 Fukayanaka, Ayase-City, 252-1195, Japan

Printed in Japan

http://www.gastec.co.jp/

Telephone +81-467-79-3910 FAX +81-467-79-3979

Advertising