Revision Date: 08/02/02

LabPro Technical Manual

20

period = time (in milliseconds) to complete one cycle


The DC output voltage is set by the equation:


V

out

= 2.4mV*amplitude – 1.2mV*offset


The command

401,1,1024,1024,0


will output a 1.25V signal. To turn off the analog output:

401,0,0,0,0


When the analog output is off, the Vin line (pin 1) of CH4 line may be used as a

±

5 volt analog input.

Note that this is different from the other three analog input channels.