Command 2001

Direct Output to Digital-Out Port

This command outputs data to the digital output port during a sampling run, thus giving the user
interactive control of some types of hardware using LabPro digital output lines. Please note the following:

•

One to sixteen data points may be output. If more than one point is sent, all the points will be sent
out at about 200

µ

sec intervals before the next command is parsed. This allows the LabPro to

clock data into a latch with a single command.

•

The output data goes to the DIG/SONIC channels. Therefore, sending this command while the
sonic port is being sampled will corrupt the data.

•

Sending this command does not stop the sampling, and it should not affect the analog channels.

•

Syntax: {2001,data1, data2, data3, . . ., dataN}

Parameter List:
data1. . .dataN

- Data values must be between 0-15. For values outside this range, behavior is

undefined.

Return values: No information is returned. The output will be effective when sampling begins.

Example 1:Turn off all lines.
Computer

Calculator

s{2001,0}

:Send({2001,0})

Example 2:Turn on all lines.
Computer

Calculator

s{2001,255}

:Send({2001,255})

Example 3:Turn on first line of port1.
Computer

Calculator

s{2001,1}

:Send({2001,1})

Example 4:Turn on first line of port2.
Computer

Calculator

s{2001,16}

:Send({2001,16})