LabPro Technical Manual

38

LabPro will collect and store a sample every 10 seconds. The recorded time for each sample will be 10
seconds. The trigger event (signal rising through 1.0 Volts) occurs 1.5 seconds after the previous sample,
so a sample collected at the trigger point is taken and stored with a recorded time of 1.5. The next sample
is taken 10 seconds after the trigger sample, not 8.5 seconds later as would have happened if the internal
sample clock had not been reset.

The Record Time returned (around the trigger point) will be the list: {...10,10,10,1.5,10,10,...}.