Index mark adjusting ring one scale division – Brookfield DV-III Rheometer User Manual

Page 29

Advertising
background image

- 29 -

When you are satisfied the pins are just hitting (by observing acceptable % readings), make a pencil mark
on the adjustment ring directly under the index mark on the pivot housing (Figure A6).

Figure A6

Turn the adjustment ring to the left (clockwise as you look down on the instrument) exactly the width
of one division from the pencil mark you have just made. This will separate the pins by 0.0005" (0.013
mm). This is a very important step because if it is not done, the pins may touch constantly and may cause
wear and damage to the cone, cup and instrument. The Rheometer is now mechanically set and ready
for sample measurement (Figure A7).

Figure A7

Adjusting Ring

Index Mark

Make a pencil

mark directly

under the

index mark

Make a Pencil Mark

Directly Under the

Index Mark

Index
Mark

Adjustment

Ring

Index Mark

Adjusting Ring

One Scale

Division

Index Mark

One Scale

Division

Turn Adjustment Ring Clockwise a

Distance Equal to One Scale Division

Adjustment

Ring