Brookfield DV-III Rheometer User Manual

Page 63

Advertising
background image

- 63 -

2. Spindle jammed below zero

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair

Display Reading Over 100

1. Overrange “EEE” (in %, cP, and SS)

• Change spindle and/or speed

Rheometer Will Not Return to Zero

1. Pivot point or jewel bearing faulty

• Perform calibration check, Section 3.3.9 of "More Solutions to Sticky Problems"
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair

Display Reading Will Not Stabilize

1. Check for erratic spindle rotation

• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair

2. Bent spindle or spindle coupling

• Replace Spindle
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair

3. Temperature fluctuation in sample fluid

• Use temperature bath control
• Check probe performance in known fluid (ice water)

4. Special characteristic of sample fluid

• Refer to Appendix C

Recorder Pen Moves in Wrong Direction

1. Output polarity reversed

• Reverse leads

No Recorder Response

1. Rheometer is at zero reading

2. Recorder is off

3. Output cable leads shorted

4. Recorder is in standby

5. Range setting is incorrect.

Advertising