- 47 -

3)

Select a viscosity standard fluid that will give viscosity readings between 10%
and 00% of full scale range. Refer to Appendix B for viscosity ranges of cone
spindles. Do not use a silicone viscosity standard fluid with a viscosity value
greater than 5000 cP with a Cone/Plate Viscometer. Brookfield offers a complete
range of mineral oil viscosity standards suitable for use with Cone/Plate Viscom-
eters as shown in Table E2. Consult with Brookfield or an authorized dealer to
determine which fluid is appropriate.

Table E2

MINERAL OIL VISCOSITY STANDARD FLUIDS

BEL Part No.

Viscosity (cP) 25

°

C

B31

31

B210

210

B750

750

B1400

1,400

B2000

2,000

B11000

11,000

B20000

20,000

B80000

80,000

B200000

200,000

B420000

420,000

Interpretation of Calibration Test Results:

When verifying the calibration of the DV-III, the instrument and viscosity standard fluid error must
be combined to calculate the total allowable error.

The DV-III is accurate to (

+

/-) 1% of any full scale spindle/speed viscosity range.

Brookfield Viscosity Standards Fluids are accurate to (

+

/-) 1% of their stated value.

Example:

Calculate the acceptable range of viscosity using RVDV-III with RV-3 Spindle at
2 RPM; Brookfield Standard Fluid 12,500 with a viscosity of 12,257 cP at 25

°

C:

1) Calculate full scale viscosity range using the equation:

Full Scale Viscosity Range [cP] = TK SMC

10,000
RPM

*

*

Where: TK

=

1.0 from Table D2

SMC

=

10 from Table D1

Full Scale Viscosity Range = 1 10 10,000

*

*

2

= 50,000 cP

The viscosity is accurate to (

+

/-) 500 cP (which is 1% of 50,000)

2) The viscosity standard fluid is 12,257 cP. Its accuracy is (

+

/-)1% of 12,257 or

(

+

/-)122.57 cP.