4 vacuum module v-802 lab vac, Vacuum module v-802 lab vac – BUCHI Vacuum Pump V-710 User Manual

Page 24

Advertising
background image

4 Description of function

24

V-700/710 Operation Manual, Version E

4 .4

Vacuum Module V-802 Lab Vac

The Vacuum Pump V-700/710 can be used together with the Vacuum Module V-802 LabVac as a
laboratory vacuum system. If the pump is used without a Vacuum Module V-802 as a laboratory
vacuum system it will automatically reduce the suction power after one hour of operation and start
working in the ECO

2

Mode (see also chapter 6.1.5).

1

2

3

4

Fig. 4.11: Front view LabVac

NOTE
The back view of the LabVac as the same as for the EasyVac, see Fig. 4.8.

1

2

a Pump not running

b Pump running

Fig. 4.12: Pressure course in LabVac mode

a Start
By pressing the START button the pump starts evacuating the system. When a constant vacuum is
reached the pump stops. When the pressure rises over the hysteresis at 50 mbar the pump starts to
operate again until a constant vacuum is reached.

Continuous
By pressing the START button for more than 2 seconds the pump enters the Continuous mode,
where the pump runs constantly.

b Stop
When pressing the STOP button the pump stops and the system is aerated by means of the valve at
the controller.