7 Maintenance and repairs

44

V-700/710 Operation Manual, Version E

• Remove the two rubber parts at the valve in-

and outlet.

• Loosen the black tube clips GL-14 between

the pump heads.

• Turn the pump on its side.

• By means of the allen key in the pump

housing (for the position of the key, see Fig.
4.2) loosen the four screws at the glass/PEEK
head.

• Remove the metal cover and the pump head.

• Use the membrane key to loosen the

membrane by unscrewing it counterclock-
wise.

NOTE
Handle the diaphragm with care and make sure
it is not damaged when removing and mounting
it.