5 Putting into operation

21

K-355 Operation Manual, Version G

5 .3 .3

Storage tank connection

To connect the storage tanks, proceed as follows:

Cut the solaflex tube into pieces of the appropriate length (use an acid resistant tube, if acid is

•

dosed).
Insert a PTFE suction tube into the solaflex tube.

•

Push an EPDM sealing ring over the solaflex tube.

•

Now fasten the tubes to the tank with the red screw cover.

•

Connect the tubes at the appropriate connections at the instrument side and secure them with the

•

clamps.

The “H

2

O” storage tank may be used to feed both the steam generator pump as well as the sample

tube pump. For this purpose the Y hose coupling can be used.

Fig. 5.3: Tank connection

All pumps are self-priming, no overpressure is necessary at the tanks.

ATTENTION
To ensure proper system functioning, consider the following:

Use only distilled water for the H

•

2

O storage tank to keep the steam generator maintenance-free.

Make sure that the tanks are connected correctly. If the wrong tank (e. g. reagent tank containing

•

NaOH) is connected to the pump labelled „H

2

O steam“, the steam generator will be damaged.