6 .2 setting the distillation time, 6 .3 setting the steam power, 6 .4 program storage – BUCHI Distillation Unit K-355 User Manual

Page 23

Advertising
background image

6 Operation

23

K-355 Operation Manual, Version G

6 .2

Setting the distillation time

To set the distillation time, proceed as follows:

Press “TIME”, the digit for minutes starts to flash.

•

Press “TIME” again, the digit for seconds starts to flash.

•

Set the distillation time by pressing “UP” and “DOWN”.

•

Start the distillation by pressing “START”.

•
Digits stop flashing after a certain time or after starting the distillation.
When pressing “UP” and “DOWN”, the time is first counted slowly, then faster.
The distillation time can be set in a range of 1 second to 99 minutes 59 seconds.

6 .3

Setting the steam power

To set the steam power, proceed as follows:

Press “STEAM”, the digit starts to flash.

•

Set the steam power by pressing “UP” and “DOWN”.

•
Digits stop flashing after a certain time or after starting the distillation.
When pressing “UP” and “DOWN”, the steam power is first counted slowly, then faster.
The steam power can be set in a range of 30 to 100%.

6 .4

Program storage

The program storage function can be used optionally. Up to 9 programs can be stored. In addition the
last used program is stored.

6 .4 .1

Storing a program

To store a program, proceed as follows:

Select a program with “UP” and “DOWN”. The steam-display shows P and the program number.

•

The program settings are shown after releasing the “UP”/”DOWN” button.
Press “STEAM” and set the steam power by pressing “UP” and “DOWN”.

•

Press “TIME”, the digit for minutes starts to flash.

•

Press “TIME” again, the digit for seconds starts to flash.

•

Set the distillation time by pressing “UP” and “DOWN”.

•

Press “REAGENT 1” until you reach the required acid or water dosing volume for storage.

•

Press “REAGENT 2” until you reach the required sodium hydroxide dosing volume for storage.

•

Press “START” to store the program.

•

6 .4 .2

Selecting a stored program

Select a stored program by pressing “UP” and “DOWN”. The steam-display shows “P” and the
program number. The program settings are shown after releasing the “UP”/”DOWN” button.

NOTE
The program settings are changed before a new distillation is started. The old program settings will
be overwritten with the new settings.