6 Operation

24

K-355 Operation Manual, Version G

6 .5

Steam generator status

6 .5 .1

Heating

A constant colon (:) and an “H” in the STEAM-display indicates, that the steam generator is heating.
A flashing colon (:) and a signal sound for 1 second indicates that the steam generator is ready.

6 .5 .2

Standby mode (after 30 minutes)

The TIME display and STEAM display show (-----).
Press “START” to cancel the standby mode.

6 .6

Switching ON/OFF checklist

To switch on and preheat the instrument, proceed as follows:

Check the filling level of the supply tanks.

•

Open the main water tap.

•

Switch on the instrument and wait until the steam generator is ready.

•

Insert an empty sample tube and an empty receiving vessel.

•

Close the protective door.

•

Set the distillation time to 2 minutes (see section “Setting the distillation time”) and the steam to

•

100% (see section “Setting the steam power”).
Start the distillation by pressing “START”.

•

To switch off the instrument, proceed as follows:

Switch off the instrument.

•

Close the main water tap.

•

Rinse the reagent 1 pump with distilled water when it is used for the additon of strong acids.

•

6 .7

Carrying out a distillation

To carry out a distillation, proceed as follows:

Switch on the instrument and wait until the steam generator is ready.

•

Insert the sample tube containing the sample.

•

Prepare a conical flask or a similar receptacle with a volume of about 250 ml as receiving vessel.

•

If required, select suitable chemicals in accordance with the method and the substance you deter-

•

mine.
Add reagents as required.

•

Set the required distillation time or select a program.

•

Start the distillation by pressing “START”.

•
For Kjeldahl distillation note the following:

Fill boric acid solution of 4% and 2-3 drops of Sher indicator into the receiving vessel. Alternative:

•

Select suitable chemicals for the back titration.
Add water to dilute the sample by pressing “REAGENT 1”. The dilution with water prevents a too

•

violent reaction when the sodium hydroxide solution is subsequently added. The sample must be
diluted in a ratio of approximately 1 part of acid to 2 parts of water.
Add sodium hydroxide solution by pressing “REAGENT 2”: Recommended alkali concentration:

•

32%. Add alkali to the sample in a ratio of about 1 part of acid to 3 parts of sodium hydroxide