4 Description of function

21

NIRFlex N-500 Operation Manual, Version F

4 .2 .1

Transflectance mode

Translucent and opaque liquids can be analyzed via transflectance mode. The light penetrates the
liquid, is diffusely reflected by the reference plate and passes the sample a second time. The trans-
flected rays contain the spectral information of the sample.

Factors that can influence the measurement:

Not enough sample material available

•

Air bubbles in the measurement path and under the transflectance cover

•

Sample not homogeneous or representative

•

Temperature of sample material

•

Sample cup or transflectance cover inadequate (e.g. cup material and thickness, blind spots, dirt

•

etc.)

NIRFlex Solids:

Diffuse reflecting / reference plate

Small, defined gap for liquid samples

Detector

NIR light

Sample cup / petri dish

NIRFlex Fiber Optic Solids / SMA:

Diffuse reflecting / reference plate

Small gap for liquid samples

Fiber optics (example)