7 .5 laser unit replacement, Warning, Notice – BUCHI NIRFlex N-500 User Manual

Page 73

Advertising
background image

7 Maintenance and repairs

73

NIRFlex N-500 Operation Manual, Version F

7 .5

Laser unit replacement

The LASER unit needs to be replaced at failure. This action can be performed by a trained operator or
(preferably) by a service technician.

!

WARNING

Risk electric shock and burns by remaining high-voltage load at the LASER module

Do not disconnect the white plug from the laser power supply

•

White plug at laser power supply

NOTICE

Risk of optical misalignment and damage at the LASER module.

Do not touch the optical surfaces of the module

•

Handle the module with care i.e. do not expose the module to heavy impacts

•

Handle the module at the base plate only

•

To replace the laser, proceed as follows:

Loosen the two screws on the left rear of the

•

instrument approximately 3-5 turns.

Remove the cover.

•

For easy release of the laser housing it is recom-
mended to slightly tilt the housing and then pull
it horizontally away from the instrument. There
is a built-in sensor to record the opening of the
housing.

Advertising