7 Maintenance and repairs

70

NIRFlex N-500 Operation Manual, Version F

!

CAUTION

Risk of minor burns by hot lamp module.

Switch off the device first and unplug the mains cable

•

Do not touch hot parts or surfaces

•

Let the lamp module cool down safely

•

Do not touch the bulb glass

•

To replace a defective lamp module, proceed as follows:

• Switch off the instrument and unplug the mains cable. Let it

cool down safely!
Loosen the two screws on the left rear of the instrument

•

approximately 3-5 turns.
Remove the cover.

•

For easy release of the housing slightly tilt the back part away
from the screws. Then pull out the housing towards the rear side
of the instrument. There is a built-in sensor to record the opening
of the housing.

A safety tack ensures that the lamp remains in

the correct position.

Slide the safety tack to the upright position (counterclock-

•

wise).
Pull the metallic housing with the primary lamp out of the

•

lamp unit.
Exchange the primary lamp source with a new lamp by

•

removing it from the gray plug and connect the new primary
lamp source to this plug.

Insert the new lamp into the lamp unit and slide the safety

•

tack back to the horizontal position until a metallic click sound
is audible, which indicates the right position. Make sure that
the gray plastic plug is in a vertical position.

Secure the housing of the lamp unit with the two screws on

•

the back side of the unit.

Replacing the lamp module