Checkline TI-UMX2 User Manual
Page 20
 
– 20 –
6.0 seleCTInG THe MeasureMenT Mode
6.1 Which mode & transducer do I use for my application?
High penetration plastics and castings
The most common mode for these types of applications is pulse-echo. The 
TI-UMX2hasbeenoptimizedforcastmaterials.Castironapplicationsrequire1-
5MHzfrequencies,andcastaluminumrequireshigherfrequenciesfrom7-10MHz
frequency. Plastics typically require lower frequencies depending on the thickness 
and make-up of the material. Larger diameters offer greater penetration power 
becauseofthecrystalsize,fordificulttomeasurematerials.
Corrosion & Pit Detection in steel and cast materials
Use pulse-echo mode whenever attempting to locate pits and flaws. Typically a 
5MHztransducer,orhigher,willbeusedforthesetypesofapplications.Uselow
frequencies for greater penetration and use higher frequencies for better resolution.
Measuring Material & Coatings
The pulse-echo coating mode should be used when both material and coating 
thicknesses are required, while still requiring the ability to detect flaws and pits. A 
special coating style transducer is required for use in this mode. There are a variety 
of coating transducers in various frequencies available from Dakota. If a dual 
element transducer is selected with the TI-UMX2 kit, it will be a coating enabled 
version.
Thru Paint & Coatings
Often times, users will be faced with applications where the material will be coated 
with paint or some other type of epoxy material. Since the velocity of the coating 
is approximately 2.5 times slower than that of steel, pulse-echo mode will induce 
error if the coating or paint is not completely removed. By using echo-echo mode, 
the user is able to successfully measure through both, the coating and steel, and 
completely eliminate the thickness of the paint or coating. Therefore, the steel can 
be measured without having to remove the coating prior to measuring. Users will 
often use pulse-echo mode and echo-echo mode in conjunction when performing 
inspections on coated materials.
Thru coating measurements require special high damped transducers. The most 
commontransducersarethe3.5,5,and7.5MHzhidampedtransducers.These
transducers are suitable for use in both pulse-echo and echo-echo modes. This 
conveniently enables the user to accurately measure overall material thickness using 
the thru Coating mode, and then conveniently switch to pit detection mode without 
changingtransducers.The_”5MHzHidampedtransduceristhemostcommonly
used transducer for standard thru coating applications.
– 37 –
Selecting the Auto Increment Direction
The Auto Increment feature gives the user the ability to specify which direction to 
advance the cursor after storing a reading.
saVInG GraPHICs: The TI-UMX2 provides the user with the ability to save 
a snapshot of the display screen and all the current settings of the TI-UMX2 
with every reading, or just save the reading only. Saving the graphics might be 
advantageous to the user for reporting purposes, and additional confidence in the 
measurements, as a snapshot of the actual A-Scan waveform will be recorded for 
every individual reading. This offers the user the ability to review and confirm the 
validity of the measurements, based on the waveforms. Additionally, enabling this 
feature will store all the current TI-UMX2 setup parameters for each measurement.
11.3 Creating a new Grid or sequential log (file)
ThissectionwillgothroughthesteptocreateaGrid/LogileandsaveitonyourPC,as
follows:
noTe: This section assumes 
that DataComm has already been 
installed on your PC, has been 
executed, and ready to go.
1. A log file can be created two
different ways; Click the icon 
buttontoinitiatethe‘New
Log’ dialog box, or click 
FILE, followed by clicking 
NEW.
2. Clickinthe‘whitespace’ields,oronthedownarrowsnexttotheieldstoenterthe
options for the file template, as illustrated above.
3. When all the information has
been entered into the fields, 
click the OK button to create 
and display the log file.
4. The log file can be saved two
different ways; Click the icon 
buttontoinitiatethe‘SAVE’
dialog box, or click FILE, 
followed by clicking SAVE.
5. Specify the path to the folder
to store the log file in, give the 
log file a unique name, and click save to store the file on your PC.