Graphics

Advanced models only

Reset restores factory settings and returns
the Gage to a known, out-of-the-box
condition. It is handy when you want to “start all over”

The following occurs:

- all batches are closed and stored measurements

are erased.

- calibration adjustments are cleared and returned

to the Gage’s factory calibration settings.

- Gates are returned to default settings.

- menu setting are returned to the following:

Memory = OFF
Graphics = OFF

(Advanced models only)

A more thorough Reset can be performed by holding
the (+) button when the Gage is powered down until
the Reset symbol appears. It performs the same
function as a menu Reset with addition of Units =
microns, and Language = English.

Setup Menu

Setup Menu

Setup

Reset

The right hand side of the screen
can be used to display a graphical
representation of the ultrasonic
pulse as it passes through the
coating system.

Press the (+) to toggle Graphics On/Off

(shortcut is active when Memory is off)

As the probe is depressed and the ultrasonic pulse
travels through the coating system, the pulse
encounters changes in density at the interfaces

Shortcut: