Checkline DFX-7 User Manual

Page 24

Advertising
background image

Dakota Ultrasonics

20

purpose of the design, was to provide the user with an efficient way to make
adjustments on the fly during inspections. The following procedure outlines the steps
to navigate and make adjustments as follows:

Hot Menu

Gate1

1) Press the MEAS key once to activate the Hot Menu. Press the MEAS key

multiple times to advance the cell right, and the ESC key multiple times to

advance the cell left, until the WIDTH1 cell is highlighted.

2) Press the LEFT and RIGHT arrow keys to toggle the value or status.

Note: This is a coarse adjustment. It’s the fastest and easiest way to

increment, decrement, or toggle the status, however it does so in coarse steps.

Set Detect

1) Press the UP and DOWN arrow keys until DETECT is highlighted.

2) Press the LEFT and RIGHT arrow keys to toggle the ABOVE or BELOW.

Note: Above will trigger the alarm when a signal meets or exceeds the height

of the gate, and below will detect on anything that meets or is below the height

of the gate.

3) Alternatively, press the ENTER key to display the Digits Edit Box.

4) Press the UP and DOWN arrow keys to scroll the highlighted value.

5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

Advertising