Checkline DFX-7 User Manual

Page 68

Advertising
background image

Dakota Ultrasonics

64

2) Press the MEAS key multiple times to move right, and the ESC key multiple

times to move left until the DAC BOX: cell is highlighted. Now that the DAC

BOX: cell is highlighted, the cursor can be controlled by moving the LEFT

and RIGHT arrow keys.

3) Couple the transducer to the first calibration standard, and peak up on the

reflector, adjusting the gain to keep the signal height within the boundaries

of the waveform display area, and at a suitable height to allow for

attenuation at greater depths in the overall range to be tested.

4) Press the LEFT and RIGHT arrow keys to move the cursor over the

waveform to be used as the first point in the DAC curve, and press the

ENTER key to set the first point. Illustrated in ex.1 above.

5) Repeat steps 3 & 4 to set multiple points as is necessary. Illustrated in ex.2

above.

Note: The DAC curve will be displayed after 3 points have been set/recorded,

and will continue to be redrawn/updated for each additional point added,

with a maximum of 8 points allowed.

6) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the DAC menu is highlighted and displaying the submenu items.