Network Configuration 33

•

•

•

•

•

•

•

Status Error code 3 (out of tolerance)

•

2: TARE CMD. Write a #2 to the command register to ZERO the net weight. The status

register will read 0 if this command succeeds:

•

Status Error code 1 (motion)

•

Status Error code 2 (A/D error)

•

4: WRITE NONVOL CMD. Write a #4 to the command register to save parameters in

non-volatile memory.

•

No Error Codes

•

0x64 (100 decimal): CAL LOW CMD. Write a 0x64 hex to the command register to

perform the low step of a traditional calibration. The status register will read 0 if this

command succeeds:

•

Calibration_Fail 1

•

Calibration_Fail_Motion 3

•

Calibration_Fail_Adc_Error 4

•

0x65 (101decimal): CAL HIGH CMD. Write a 0x65 hex to the command register to

perform the high step of a traditional calibration.

•

Calibration_Fail 1

•

Calibration_Fail_Motion 3

•

Calibration_Fail_Adc_Error 4

•

HardcalFailCounts 8: not enough counts between hard cal hi and hard cal lo

•

0x66 (102 decimal): C2 CAL CMD. Write a 0x66 hex to the command register to

perform a C2 calibration.

•

Calibration_Fail 1

•

Calibration_Fail_Motion 3

•

Calibration_Fail_Adc_Error 4

•

Calibration_Fail_Noc2 5

•

Calibration_Fail_C2capeq 6

•

Calibration_Fail_C2clones 7

•

0x1000 (4096 decimal): WRITE INTEGER CMD. Set the value of an integer

parameter. Write 0x1000 in the command register 0, the parameter ID number in

registers 2, 3 and the desired value in registers 4, 5.

•

No Error Codes

•

0x1001 (4097 decimal): WRITE FLOAT CMD. Set the value of a floating point

parameter. Write 0x1001 in the command register 0, the parameter ID number in

registers 2, 3 and the desired float value in registers 4, 5.

•

Failure 1