Network Configuration 37

The 4-20mA parameters can be configured through the web interface or the display panel.

From the front display: select the configuration menu then UP or DOWN to
communications, enter; UP or DOWN to 4-20mA, enter; UP or DOWN to setting you
want for Source; enter the value for 4mA; enter the value for 20mA or mA output (read
only). Press enter to select and UP or DOWN arrows to enter value.From the Website,
select the Communications Menu and then

Step 1. Select the source for the 4 to 20mA output, default is gross

a. Gross.

b. Net

c. Count

d. Manual

Step 2. Set the 4mA value

Step 3. Set he 20mA value

Step 4. Save Parameters

NOTE

Setting the weigh source to manual allows the mA output setting to be set to a fixed
output. This is used to test the circuit level to the host.