Establishing zero – Holtgreven GSE-355 I.S. User Manual

Page 98

Advertising
background image

89

Establishing Zero

The Model 350/355 provides five methods for obtaining a zero (no load) calibration reference, First Zero, Last Zero,
False Zero, Only Zero, and Cal Reset.

To select a calibration method (350):

1. Press



+



to display the calibration prompt.

2. Press



О



О

2

О

1

to access the Calibration Mode.

3. Press



to scroll through the five selections.

4. Press

1

to establish zero.

To select a calibration method (355):

1. Press



!

to display the calibration prompt.

2. Press



(

to access the Calibration Mode.

3. Press

!

to scroll through the five selections.

4. Press

(

to establish zero.

F

IRST

Z

ERO

The most common zeroing procedure, First Zero is used to establish a new zero (no load) calibration reference before
proceeding to span the Model 350/355. Use this method for first-time calibration and complete recalibration.

First Zero Calibration Method Example (350):

1. From the Weigh Mode, press



+



.

DISPLAY READS X Setup

2. Press



О



О

2

О

1

.

DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ -0.26

3. Remove any load on the scale.

DISPLAY READS X First ~ Zero? ~ -0.42

4. Press

1

to establish zero.

DISPLAY READS X 0.00

5. Pause for motion delay.



Press



or

during calibration to back up one step in the procedure.