IDS440 Users Manual version 4.2

Set Operation Parameters

21

Set Operation Parameters

This section describes the operation parameters of the IDS440 and how their settings

affect the operation of the weight indicator.

Access Operation Parameters and Parameter Functions

The parameters are accessed by holding the CLEAR key down and then pressing the

ENTER key. The numeric display will prompt “CFG xx” where xx is the currently

selected parameter. Enter the parameter number to be modified and then press the
ENTER key. NOTE: the UNITS key (

↓) and the PRINT key (↑) can be used to scroll

the CFG parameter number. Press the UNITS key to increment the CFG number, press

the PRINT key to decrement the CFG number.
To exit configuration and store changes, press the CLEAR key when “CFG xx” is being

displayed.
To exit configuration and NOT store changes, enter 999 when “CFG xx” is being

displayed. The display prompts “Abort”. Press the ENTER key to abort (exit without

saving changes) or press the CLEAR key if you do not want to abort.

Operation Parameters Reference Tables

The reference tables provide an overview of the operation parameters. The sections

following the reference tables describe the setup parameters in detail and the procedures

for modifying them. Use the ‘Field Settings’ column of the reference tables to document

any changes to the setup parameters.
Reference Table column headings descriptions:

No.: Use this number to access the parameter to be modified. For example to modify

the Digital Filter enter CFG 2.

Parameter Name: The name of the parameter.

Default: The value set at the factory. The Parameter Function “Initialize Operation

Parameters to Factory Defaults” sets the parameter to this value.

Max Value: The maximum value for numeric parameters, the maximum number of

characters for alphanumeric parameters.

Pass: The IDS440 has 4 password levels. Level 1 requires entering password-1

before accessing the parameter. Level 2 requires password-2. Level 3 requires

password-2 and the Lock Switch to be in the unlocked position. Level 4 inhibits Scale

Basic functions from being uploaded to a PC.

Field Setting: Record any changes to the parameters in the Field Setting column.