IDS440 Users Manual version 4.2

Parameter Functions

46

Parameter 76 Debug mode for Event monitors and Scale Basic

The Debug mode is used to ‘debug’ event monitors and scale basic functions. Enter a 1

to activate debug mode, enter a 0 to disable debug mode. When debug mode is enabled

Scale Basic functions ‘single step’. The subroutine stack level and the Function number

are displayed when entering a function. Example: “0 F12”. Press the ENTER key to

begin execution of the function. The instruction address and instruction code are

displayed. Example: “21 255”. Continue to press the ENTER key to step thru the

program. NOTE: the first step displayed for all scale basic functions is a GOTO (244).

Function 78 Initialize ID memory

Use function 78 to initialize id memory after it has been corrupted. The ID memory

becomes corrupted when power is lost (when battery backup is disabled) or when the

battery is allowed to discharge below 2 volts.

Parameter 79 Enable Multi-point Linearization

Enter a to 1 to enable multi-point Linearization, enter a to 0 to disable multi-point

Linearization. The IDS 440 uses the calibration factors set in function 80 below when

multi-point Linearization is enabled.

Function 80 Calibrate - Multi-point Linearization

Use function 80 to calibrate the scale using up to 5 calibration points to 'linearize' the

load cell. Set the Scale Operation Parameters before using function 80. Best results are

obtained when the IDS 440 and load cells have reached thermal equilibrium (power

applied for approximately 30 to 120 minutes depending on load cell size). NOTE:

function 80 automatically adjusts parameter 19 (deadload offset).
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.

2. Enter CFG 80. The IDS 440 prompts "dEAdld". Remove all weight from the scale.

Press the ENTER key to obtain the deadload weight. Press the Units key (

↓) to

continue without reading a new deadload weight.

3. The IDS 440 ‘s display alternates between "CAL 1" and the weight on the scale.

Place the first calibration weight on the scale. Enter the calibration weight using the

numeric keypad. Press the ENTER key. The IDS 440 reads the calibration weight.

4. The IDS 440 ‘s display alternates between "CAL 2" and the weight on the scale.

Repeat step 3. The IDS 440 will accept up to 5 calibration points. Press the CLEAR

key to complete calibration if less than 5 calibration points are used.

5. The IDS 440 prompts "CFG 80". Press the CLEAR key to exit the configure mode