background image

IDS440 Users Manual version 4.2

53

Hardware and Wiring

This section describes installation and wiring information for the interface ports. There

are two bi-directional serial ports, one parallel port with TTL inputs and outputs, and one

load cell input port. The serial ports are used to interface to a printer and to a host device

or PC for continuous output. The parallel port is used to interface to parallel printers and

for remote switch input and relay control using the TTL I/O. The load cell input is used

to interface to the scale platform.

Load Cell Connector - TB3

NOTE: JP1 and JP2 on the circuit board can be used to disable remote sensing.

Load Cell - TB3

Pin # Signal Name

1

+ Signal

2

- Signal

3

+ Excitation

4

+ Sense

5

- Sense

6

- Excitation

7

Analog ground

JP1

JP2

TB3

Remote Sense Jumpers - Remote sense DISABLED

Load Cell Terminal Block

Place jumpers as shown to DISABLE remote sense. Move jumpers up to
enable remote sense.