Appendix c: troubleshooting the load disc system – Kistler-Morse KM LD3xi Load Cell User Manual

Page 31

Advertising
background image

Appendix C. Troubleshooting

C-1

Appendix C: Troubleshooting the

Load Disc System

This section describes an output check and some common
problems. For each problem, one or more possible explana-
tions are listed. For each explanation, suggested solutions
are provided.

1.

Mesure the output of the load cell using procedure from
page 2-2.

2.

Verify the output to be between 0mV and +/- 1mV, stable.

3.

Repeat Steps 1 and 2 for each LD3xi.

4.

If the load cells are installed under the vessel, verify
stability of each load cell.

Note: The “no-load”
condition is when the
Load Disc stands
alone without any
weight applied.

Functional Check: Measuring Output

(while wired to Signal Processor)

The following will be true between 0 °F and 100 °F:

1.

Measure between the disconnected excitation wires and verify the
resistance to be 700 ohms +/- 15 ohms, with a stable reading.

2.

Measure between the disconnected output wires and verify the
resistance to be 700 ohms +/- 15 ohms, with a stable reading.

Functional Check:Measuring Resistance

Note: When using the 61-6036-01 trim
box, and a sensor fails, the sensor must
be replaced. When the sensor wires are
removed from the junction box, the
jumper must be soldered back in place.