25

9-5 Program setting mode

The nine programs can be saved in the program memory built into the pipette (PROG
01 to 09). By saving a frequently used mode or volume for operation beforehand,
these setting can easily be read out from the next use. Select and set the mode or
volume to be saved before saving the program setting.

Saving the program setting
Set the pipette to your preferred settings.

Example: When saving the AUTO mode, dispensing volume 200μL, buzzer ON and

blowout ON

(The display is example.)

[ 1 ] Press and hold the Down key to enter

program setting saving mode.

[ 2 ] Press the Up or Down key to select program

number to be save from 1 to 9.


[ 3 ] Press the Enter key to save the setting.

Once these have been saved a buzzer will

sound once.

Reading out the program setting

The set program can be read out at system setting mode (SYS).
(Refer to “9-4 system setting mode (SYS)” for details)

9-6 Reverse operation (Dispensing liquid that tends to remain in the tip)

When you would like to accurately dispense a highly viscous liquid that has a
tendency to remain in the tip, we recommend using the reverse operation. By
aspirating a large amount of the liquid beforehand, the reverse operation enables the
correction of the amount of liquid remaining in the tip.

To enable reverse operation, set the setting of at system setting mode (SYS).

(Refer to “9-4 System setting mode (SYS)”.)