3 dispensing in a unit of weight (in mg unit) – A&D Weighing MPA-1200 User Manual

Page 33

Advertising
background image

31

[ 3 ] Press the Up or Down key to select volume

calibration data resetting mode.



[ 4 ] Press the Enter key to enter volume

calibration data resetting mode.


[ 5 ] Press the Up or Down key to select .

(

means YES and

means NO.)

[ 6 ] Press the Enter key to reset volume

calibration data.



[ 7 ] The Notice mark is turned off to indicate

that volume calibration data have reverted back
to the factory default.

11-3 Dispensing in a unit of weight (in mg unit)

Dispensing of a liquid can be performed by weight (mg) instead of volume. (mg unit function)
This function is useful when you handle a liquid that needs to be managed by weight,
such as a diluted solution of a solid or powder. Although the density of a liquid can
vary depending on the sample type and concentration, by weighing the dispensed
amount with an electronic balance and inputting the result into the pipette, it becomes
possible to easily dispense the liquid in a unit of weight (mg).

Selecting the mg unit

The unit (volume: μL / weight: mg) for pipetting can be toggled by the following method.
When the mg unit is selected, the Notice mark blinks and the μL unit is turned off.

* When the unit of weight (mg unit) is selected, perform weight calibration by the

dispensing amount to be used. (The weight calibration data reverts back to the
factory default once the unit is switched toμL.)




Volume calibration data

resetting mode