Calibration – ALTANA Haze-Gloss Manual User Manual
Page 13

• Call SET UP Menu
SET UP OPERATION
Cursor: SET UP
Change: OPERATE
Continue: SET UP
5. Calibration
• Call SET UP Menu
SET UP OPERATION
Cursor: SET UP
Change: OPERATE
Continue: SET UP
5. Calibration
Calibration of the
haze-gloss is seldom
necessary due to state of the art
technology and a reference beam
principle. We recommend a calibration
interval of
≥ 2 months.
Calibration is also recommended
• if the ambient temperature differs
extremely from the normal room
temperature.
• if "Please Recalibrate" appears on the
display. (The unit automatically de-
tects lamp aging by 4 %)
set up
Calibration Procedure
Calibration Procedure
set up
Calibration of the
haze-gloss is seldom
necessary due to state of the art
technology and a reference beam
principle. We recommend a calibration
interval of
≥ 2 months.
Calibration is also recommended
• if the ambient temperature differs
extremely from the normal room
temperature.
• if "Please Recalibrate" appears on the
display. (The unit automatically de-
tects lamp aging by 4 %)