Standard Operations

5.7 Photometry

In the photometry mode, photometric measurements can be carried
out as:

•

Single wavelength measurements

•

Multi wavelength measurements

•

Time course

•

Wavelength Scans

5.7.1 Single Wavelength (absorbance, concentration and transmittance measurements)

The Single Wavelength mode can be used in three ways. For
sample measurements at a single wavelength, the instrument can
be programmed to measure the absorbance, % transmittance or
concentration of the analyte.

Absorbance measures the amount of light absorbed by the sample,
in units of Absorbance.

% transmittance measures the percent of the original light that
passes through the sample and reaches the detector.

Turn the concentration factor on to select a specific multiplier for
converting absorbance readings to concentration. In a graph of
concentration versus the absorbance, the concentration factor is
the slope of the line.