6.2 Calibrate
Turn the unit on by pressing either the menu scroll
wheel or the operate button.
Choose the menu item Calibrate from the Calibrate
menu.
Remove the protective cap. Place the instrument on
the black trap. Make sure that the instrument fits
into the indentation of the black trap.
Choose OK with the wheel and press the operate
button to perform the black trap reading.
The reading process is indicated on the display.
Choosing Cancel will abort
the calibration procedure and calibration will
not be updated.
Calibrate