ALTANA Byko-Test MP0R User Manual

Page 36

Advertising
background image

36

Normalization performed by mistake on calibration
standard instead of on uncoated specimen.

Perform normalization on uncoated specimen.

Er14

Internal instrument error: Not able to compute
master characteristic. The original master
characteristic is retained.

Repeat master calibration.
If this error occurs repeatedly:
Contact customer service.

Er15

Not able to save the master characteristic.
Probe may be defective.

Contact customer service.

Er17

Not enough readings. Thus, not able to finish a
block (applies to measurements w. fixed block
limits).

Make additional measurements until the first
block can be finished. Then generate a new final
result.

Er22

Internal instrument error. Error accessing the
measurement memory or. EEPROM.

Contact customer service.