Aneroid sphygmomanometer, Size chart, Device description and intended use – American Diagnostic Corporation (ADC) Diagnostix 75XX User Manual

Page 2: Contraindications, General warnings

Advertising
background image

ADC

®

Aneroid Sphygmomanometer

(manometer, cuff, bladder, bulb and valve)

Thank you for choosing an ADC

®

DIAGNOSTIX

TM

Aneroid Sphygmomanometer. Your new

ADC

®

DIAGNOSTIX

TM

aneroid is designed to provide unrivaled durability, and unparalleled

comfort. Every component has been carefully designed to maximize performance. This

Use, Care, & Maintenance guide refers to ADC model series; 700, 703, 705, 720, 728, 731,

732, 740, 750, 752, 778, and 788.

Device Description and Intended Use

ADC

®

aneroid sphygmomanometers are used by professional healthcare providers and in-

dividuals trained in the auscultatory blood pressure technique to determine systolic and di-

astolic blood pressure in humans.

Contraindications:

Aneroid sphygmomanometers

are contraindicated for neonate use.

Do not use with neonatal cuffs or

neonate patients. Review the size

chart (right) for proper age and

limb range usage.

General Warnings

A warning statement in this manual identifies a condition or practice which,

if not corrected or discontinued immediately could lead to patient injury, ill-

ness, or death.

WARNING: Do not allow a blood pressure cuff to remain on patient for more than 10

minutes when inflated above 10 mmHg. This may cause patient distress, disturb blood

circulation, and contribute to the injury of peripheral nerves.

WARNING: If luer lock connectors are used in the construction of tubing, there is a

possibility that they might be inadvertently connected to intra-vascular fluid systems,

allowing air to be pumped into a blood vessel. Immediately consult a physician if this occurs.

WARNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is not estab-

lished.

WARNING: If this equipment is modified, appropriate inspection and testing must be

conducted to ensure its continued safe use.

Cuff Size

Limb Range

Inches CM

Infant

7 3.5 to 5.5 9 to 14

Child

9 5.1 to 7.6 13 to 19.5

Sm. Adult

10 7.4 to 10.6 19 to 27

Adult

11 9 to 15.7 23 to 40

Lg. Adult

12 13.3 to 19.6 34 to 50

Thigh

13 15.7 to 25.9 40 to 66

Size Chart

Advertising