Measurement Procedure

1. Patient Position

The patient should sit or lie comfortably. The arm should be fully supported on a flat

surface at heart level. (If the arm’s position varies, or is not level with the heart, measure-

ment values obtained will not be consistent with the patient’s true blood pressure.) When

seated, the patient should have their back and arm supported, and their legs should not be

crossed. The patient should relax prior to measurement comfortably for five (5) minutes

and should refrain from talking or moving during measurement. Observer should view

manometer in a direct line to avoid “Parallax error”.

2. Apply the cuff

ADCUFF

TM

nylon cuffs are specially designed to promote the precisely accurate determi-

nation of blood pressure. Index and range markings ensure use of the correct cuff size.

The artery mark indicates proper cuff positioning.

Place the cuff over the bare upper arm with

the artery mark positioned directly over the

brachial artery. The bottom edge of the cuff

should be positioned approximately (1”) one

inch (2-3cm) above the antecubital fold.

Wrap the end of the cuff not containing the

bladder around the arm snugly and smoothly

and engage adhesive strips (Figure 1).

To verify a correct fit, check that the Index Line falls between the two Range Lines (Figure

2).

3. Inflate the cuff

Close the ADFLOW

TM

valve by turning thumbscrew clockwise.

Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by squeezing

bulb rapidly.

Left Arm

Tube

(Figure 1)

Index Line

Range Lines

Index Line

Artery Mark

Select Smaller Cuff

Correct Cuff Size

Select Larger Cuff

(Figure 2)