printed oval when the unit is fully deflated (Figure 3).

Should the indicator needle of the manometer rest outside of

this calibration mark, then the manometer must be re-calibrated

to within ±3 mmHg when compared to a reference device that

has been certified to national or international measurement

standards. A manometer whose indicator needle is resting out-

side of this mark, is NOT acceptable for use.

In the event that the gauge is ever in need of calibration, simply return to ADC. Damaged or

broken parts will be replaced as needed at a minimum charge. Refer to the warranty for spe-

cific details of warranty coverage.

The manufacturer recommends a calibration check every 2 years.

Cuff Cleaning and Disinfecting

NOTE: Use one or more of the following methods and allow to air dry:

• Wipe with mild detergent and water solution (1:9 solution). Rinse.

• Wipe with Enzol per manufacturer’s instructions. Rinse.

• Wipe with .5% bleach and water solution. Rinse

• Wipe with 70% isopropyl alcohol.

• Launder with mild detergent in warm water, normal wash cycle.

Remove bladder first. Cuff is compatible with 5 wash cycles.

Low Level Disinfection

Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray deter-

gent solution liberally onto cuff and use a sterile brush to agitate the detergent solution over

entire cuff surface for five minutes. Rinse continuously with distilled water for five minutes.

To disinfect, first follow the cleaning steps above, then spray cuff with 10% bleach solution

until saturated, agitate with a sterile brush over entire cuff surface for five minutes. Rinse

continuously with distilled water for five minutes. Wipe off excess water with sterile cloth

and allow cuff to air dry.

CAUTION: Do not iron cuff.

CAUTION: Do not heat or steam sterilize cuff.

NOTE: Store gauge with

valve in full exhaust position.