Defibtech AED Training Equipment User Manual

Page 9

Advertising
background image

5

DAC-515E-EN-BD

3.3 Available Scenarios

Six training scenarios (described below) are available for training exercises, and

can be changed during training as desired.

The following scenarios are available:

1. VF (Ventricular Fibrillation) that converts to a non-shockable rhythm

(normal sinus rhythm) after the 1st shock. This scenario is based

on the American Heart Association AED training scenario and

allows for a quick overview and demonstration of AED operation.

2. Non-shockable rhythm throughout.
3. Bad pads indication until the pads are disconnected and then

re-connected (simulating replacement of the pads), followed by

VF that converts after the 1st shock to a non-shockable rhythm

(normal sinus rhythm).

4. VF that converts on the 2nd shock to a non-shockable rhythm

(normal sinus rhythm).

5. Non-converting persistent VF.
6. Pads not applied to the patient. This scenario is recommended as

the Default Scenario when using the Remote Control. The AED

will prompt the student to apply pads to the patient dummy, and

the instructor can then manually select rhythm simulations (such

as NSR and VF) or any one of the above training scenarios using

the Remote Control.

Note: When the AED Trainer is turned on, it will power up in the Default

Scenario, not in the last-used scenario (see section 3.5 for instructions on setting

the Default Scenario).
Note: Training scenarios 1-5 assume that pads have already been applied to

the patient if pads are connected to the AED when the AED is turned on. For

properly sequenced exercises using these training scenarios, the student should

either apply the pads to the patient before turning the unit on, or turn the unit

on with pads disconnected, apply the pads to the patient, and then connect the

pads to the AED.

3.4 The Default Training Scenario

When a AED Trainer is first turned on, it operates in the “Default Training

Scenario”. The Default Training Scenario can be changed to any of the six available

training scenarios.
Note: New Standalone AED Trainers and newly-configured DDU-100 AEDs are

factory programmed to initially start with the Default Training Scenario set to

scenario #6.

Advertising